Data
Measurement range
0.01 to 10,000 ppb
Measurement method
Continuous sodium measurement using ISE (Sample time programmable for multi-stream version)
Accuracy / maximum measured error
±5 % of reading or ±0.1 ppb, whichever is the greater
Repeatability
±5 % of reading or ±0.1 ppb, whichever is the greater
Response time
1 to 100 ppb <4 mins;
100 to 1 ppb <4 mins for a 90 % step change
Resolution
0.001 ppb
Calibration
2-point, automatic calibration
Measurement range
0.01 to 10,000 ppb
Measurement method
Continuous sodium measurement using ISE (Sample time programmable for multi-stream version)
Accuracy / maximum measured error
±5 % of reading or ±0.1 ppb, whichever is the greater
Repeatability
±5 % of reading or ±0.1 ppb, whichever is the greater
Response time
1 to 100 ppb <4 mins;
100 to 1 ppb <4 mins for a 90 % step change
Resolution
0.001 ppb
Calibration
2-point, automatic calibration